PTA Scoring Balloon Dilatation Catheter

DKutting™

PTA Scoring Balloon Dilatation Catheter

NMPA 20223031527

Indicated to dilate lesions in the iliac, femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Technical Specifications

 

Catheter Type OTW
Guidewire Compatibility 0.018” & 0.035”
Compliance Type Non-compliant
Balloon Diameters 4.0, 5.0, 6.0, 7.0, 8.0 mm
Balloon Lengths 20, 40, 60, 80 mm
Catheter Lengths 50, 90, 130 cm
Sheath Compatibility 7F
Nominal Pressure (NOM) 8 atm
Rated Burst Pressure (RBP) 20 atm
Balloon Material Nylon
Balloon Coating Hydrophobic coating
Scoring Element Material Nitinol
Scoring Element Height 0.36 mm

 

 

0.018″ Balloon Length (mm)
Balloon Diameter (mm) 20mm 40mm 60mm 80mm
4.00mm
5.00mm
6.00mm
7.00mm
8.00mm
0.035″ Balloon Length (mm)
Balloon Diameter (mm) 20mm 40mm 60mm 80mm
4.00mm
5.00mm
6.00mm
7.00mm
8.00mm

*Same size for 50, 90 and 130cm catheter length

DKutting™ - PTA Scoring Balloon Dilatation Catheter

Technical Specifications

 

Catheter Type OTW
Guidewire Compatibility 0.018” & 0.035”
Compliance Type Non-compliant
Balloon Diameters 4.0, 5.0, 6.0, 7.0, 8.0 mm
Balloon Lengths 20, 40, 60, 80 mm
Catheter Lengths 50, 90, 130 cm
Sheath Compatibility 7F
Nominal Pressure (NOM) 8 atm
Rated Burst Pressure (RBP) 20 atm
Balloon Material Nylon
Balloon Coating Hydrophobic coating
Scoring Element Material Nitinol
Scoring Element Height 0.36 mm

 

 

DKutting™ - PTA Scoring Balloon Dilatation Catheter