| Balloon Diameter (mm) | Balloon Diameter (mm) | |||||
| 10 | 15 | 20 | 25 | 30 | 35 | |
| 2.00 | √ | √ | √ | √ | √ | √ |
| 2.25 | √ | √ | √ | √ | √ | √ |
| 2.50 | √ | √ | √ | √ | √ | √ |
| 2.75 | √ | √ | √ | √ | √ | √ |
| 3.00 | √ | √ | √ | √ | √ | √ |
| 3.50 | √ | √ | √ | √ | √ | √ |
| 4.00 | √ | √ | √ | √ | √ | √ |
- Technical Specifications
- Available Configurations
- Resources
Technical Specifications
Technical Specifications
| Catheter Type | Rx |
| Guidewire Compatibility | 0.014” |
| Compliance Type | Non-compliant |
| Balloon Diameters | 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 4.0 mm |
| Balloon Lengths | 10, 15, 20, 25, 30, 35 mm |
| Catheter Lengths | 142 cm |
| Guiding Catheter Compatibility | 5F |
| Nominal Pressure (NOM) | 10 atm |
| Rated Burst Pressure (RBP) | 16 atm |
| Balloon Material | Nylon |
| Tip Entry Profile | 0.016” |
| Drug Dose | 3.0 µg per m㎡ of the balloon surface |
| Excipient | Midchain Triglyceride |
Available Configurations